Programming Guide
9-4
Error Messages
Error Description
Cause of the error
Action
Incorrect parameter is
set for cooling.
Check and correct the
parameters.
COOL TIME OUT
The temperature
does not decrease to
the set value within
the time limit from
cooling start.
Defect in the heating
system.
Contact service
center.
OVER TEMP.
The temperature
exceeds the set value
while heating.
Defect in the heating
system.
Contact service
center.
Fiber is dirty.
Retry the splice from
preparation. Make
sure to clean the bare
portion of the fiber.
Defect in the image
processing system.
Contact service
center.
VISUAL ERROR
The image process
cannot focus on the
fiber, find the clad
line.
Optics is dirty.
See Maintenance
chapter.
OUT OF SPEC
The fiber is out of
applicable range.
Cladding diameter is
out of applicable
range.
Can not splice with
S153 Ver.2.
Содержание S153 Ver.2
Страница 2: ......
Страница 8: ......
Страница 22: ......
Страница 28: ......
Страница 40: ...External Description 5 2 Electrode V Groove Fiber Holder Fiber Clamp ...
Страница 49: ...External Description 5 11 Function Menu Function Keys Press to select the corresponding Function Menu ...
Страница 50: ......
Страница 99: ...Programming Guide 8 21 Time chart of heater parameters ...
Страница 152: ......
Страница 158: ...Option 10 6 10 4 Cleaning Brush VGC 01 It is used to remove the garbage that adheres to V groove and the fiber clamp ...
Страница 163: ...Recycling and Disposal For sales and service information contact FURUKAWA ELECTRIC CO LTD or your local representative ...