63
and retry Arc Check”. Perform another arc check to optimize the arc
power.
When executing the "Self Check” function, please use fibers that
have been stripped, cleaned and cleaved correctly.
6.7.2 Fiber Measurement
The S124 performs an auto or manual inspection of the fiber (specifically,
the clad and core offset, relative eccentricity, gap, fiber tilt and relative
cleave angle).
1st Result Screen (Bilateral measurement)
Parameter
Description
Clad Offset
Amount of clad offset between the two fibers.
Core Offset
Amount of core offset between the two fibers.
Rel. Eccentricity
Difference in eccentricity between the two fibers.
Gap
The gap between the two fibers.
Fiber Tilt
Angle at which the fibers come into the screen.
Rel. Angle
The relative cleave angle between the two fibers.
2
nd
and 3
rd
Result Screen (Right and Left fiber measurement)
Parameter
Description
Eccentricity
Eccentricity of the fibers in micrometers.
Cleave Angle
Cleave Angle of the fiber in degrees.
Clad Tilt
Angle of Clad.
Clad Width
Measurement of clad width in micrometers.
Core Tilt
Angle of Core.
Core Width
Measurement of core width in micrometers.
Focus
Focus value (%).