6.3 Fusion Splicing
1.
Ensure that the “READY” screen is displayed on the monitor.
2.
Press
to initiate the fusion splicing cycle.
3.
The S124 fusion splicer performs the following functions automatically.
To pause the S124 fusion splicer during any of these functions, press
. The message PAUSE will be displayed on the monitor. To restart
the operation, press
again.
The right and left fiber ends appear on the LCD monitor.
A cleaning arc is discharged to clean the fiber ends.
The fibers are set with a gap of about 20 µm between the ends.
The screens are zoomed up.
The fibers are inspected for axis offset and cleave condition.
The cores of the fibers are aligned on the X and Y view screens.
The electrodes discharge.
The splice is inspected.
The splice loss is estimated and displayed on the LCD monitor as
shown in the picture.