84
Always replace or clean both electrodes, even if only one electrode is
damaged.
Be sure to turn off the power switch before starting maintenance.
Never touch the electrode while the Power is on.
Longer arc duration used in dissimilar fiber splicing requires the
electrodes to be cleaned and replaced more often. Frequent electrode
maintenance is recommended for dissimilar fiber splicing programs.
1.
Loosen the screws of the holding plates, and raise the plates. The
electrode is raised together with the holding plate. Be careful not to
drop the electrodes into the machine.
2.
Carefully pull and remove the electrodes from the holding plates by
grasping the electrode knob. Make sure nothing touches the electrodes
tips.
3.
Clean or replace the electrodes, as necessary.
Screw
V-groove
Electrode
Electrode knob
Holding plate