EVO EXPERT Manual
85/105
Table 20. VANE special spindle selection for EVO L
Maximum guideline values in cP (mPa·s)
RPM / SP
V71
V72
V73
V74
V75
0.01
245K
1.04M
5.01M
50.8M
21.6M
0.3
8.18K
34.6K
167K
1.69M
721K
0.5
4.91K
20.8K
100K
1.01M
433K
0.6
4.09K
17.3K
83.5K
848K
360K
1
2.45K
10.4K
50.1K
508K
216K
1.5
1.63K
6.93K
33.4K
339K
144K
2
1.22K
5.20K
25.0K
254K
108K
2.5
982.2
4.16K
20.0K
203K
86.6K
3
818.5
3.46K
16.7K
169K
72.1K
4
613.9
2.60K
12.5K
127K
54.1K
5
491.1
2.08K
10.0K
101K
43.3K
6
409.2
1.73K
8.35K
84.8K
36.0K
10
245.5
1.04K
5.01K
50.8K
21.6K
12
204.6
867.0
4.17K
42.4K
18.0K
20
122.7
520.2
2.50K
25.4K
10.8K
30
81.85
346.8
1.67K
16.9K
7.21K
50
49.11
208.0
1.00K
10.1K
4.33K
60
40.92
173.4
835.7
8.48K
3.60K
100
24.55
104.0
501.4
5.08K
2.16K
200
12.27
52.02
250.7
2.54K
1.08K
ATTENTION:
K Indicates miles.
Example: 7.8K = 7800
M Indicates Millions
Example: 1.56M = 1560000
NOTE:
It is not recommended to work with viscosity values of less than 15% of the lower part of