EVO EXPERT Manual
33/105
To choose one of the tests profile, press the corresponding key for the test profile that is desired.
The names correspond to the symbols that there are on each of the keys on the apparatus´
keyboard (for example hitting the key ‘6 M6’ selects log M6). From there, hit the ‘ENTER’ key to
validate the option.
In the test profile recording there are two option blocks that you must to configure once the desired
test profile has been chosen. We will now explain viscometer programming and output specific
configuration for the measurement.
8.4.1.1 Viscometer programming
Once the log is chosen, the following screen will appear:
For the selection of one of the two options, scroll between the options by using the ‘
’ and ‘
’ keys
and press the ‘ENTER’ key on the one that is desired. The exit keys, ‘MEM/CLEAR’ and ‘
◄
’,
continue to fulfil their habitual functions by bringing the user to the main menu screen or the
previous screen, respectively. In the case of ‘MEM/CLEAR’, it will proceed without having saved the
changes.
On this screen, these two fields can be configured. Once they are configured, the 0N key accesses
to the ‘MEASURE CONFIGURATION’ screen, which should be filled with the main parameters desired
for the measurement, such as the spindle, the motor speed and the density of the sample.
8.4.1.1.1 TTT and TTS
These abbreviations mean:
TTT:
Time to Torque. You must set a torque value (%), at which the viscometer will
have to stop the measurement. The screen will show the obtained viscosity at
this moment in the torque. (see section 8.5)
TTS:
Time to Stop. You must set a time for the experiment and a time for the
viscometer to stop. Once the device has arrived at the determined time, the
equipment will stop and display the value of the viscosity (see section 8.5)