EVO EXPERT Manual
24/105
Use the numerical keyboard to introduce the value of the viscosity of the standard oil used for
calibration (the standard oils provided by Fungilab provide viscosity tables according to different
working temperatures). There is a field for entire numbers and other one for the decimal figures.
Once introduced the density value, press ‘ENTER’ with the ‘SAVE’ option highlighted to confirm the
modification. Next, the following screen will appear:
Once the spindle is in position in the device, press ‘ENTER’ again and the following screen will
appear:
In this screen it is necessary to introduce the time required from the moment you give the
command to start the calibration to the moment the device begins the calibration process. This time
lapse is frequently used to allow the whole of the sample and spindle to arrive at thermal stability
before starting the actual calibration.
On this screen, the field for the hours appears highlighted first. Using the ‘
’ and ‘
’ keys you can
change the number of hours. Once the right value is entered, hit ‘ENTER’ and the field associated
with the number of minutes will stay highlighted and ready to be modified using the ‘
’ and ‘
’
keys. Following this same procedure the number of seconds can be modified. When pressing the
‘ENTER’ key with the ‘SAVE’ option highlighted it will start a countdown back to zero. The following
screen can be an example of this countdown: