EVO EXPERT Manual
6/105
This arrow indicates additional information that should be
used by the user.
This symbol warns us of an operational, practical, or similar
procedure that, if it is not carried out correctly, may
irreparably damage the equipment. Do not proceed further
unless the indicated conditions are fulfilled and have been
perfectly understood.
4.
Conditions for use
-
Indoor use
-
Maximum altitude 2000 m.
-
Surrounding temperature range: from +5 to 40ºC.
-
80% maximum relative humidity for up to 31ºC and going as low as 50% of relative
humidity for up to 40ºC.
-
The power source fluctuations should not surpass
10% of the nominal voltage
-
Installation category II
-
Pollution level II
5. Maintenance
Always clean all of the parts after each use! Clean and dry the spindles and the spindle
guard well. Make sure that there is not any sample remaining left, especially in the delicate
zones like the spindle connector.
Use detergents or solvents to clean the spindles and the protector:
-
For cleaning food samples, use lukewarm water and if necessary, use soft
household detergents
-
Other solvents that generally give good results are acetone, gasoline, or any
solvent with a high percentage of alcohol
-
For the use of any other solvent, make sure that it does not corrode the spindles or
the protector. The spindles are made in AISI 316.
.
Warning
: Handle the volatile and inflammable solvents with proper cautions.
It is the user’s responsibility to establish safety conditions at work.
Regularly check the spindle’s thread and the viscometer shaft.
During the viscometer’s lifespan, the equipment might require certain check-ups to perform
as expected. For this service, please contact the local distributor.
Regular maintenance is important. As the manufacturer, we advise annual check-
ups by the technical service of your local distributor.
The viscometer is powered by the MEAN WELL GS25A12-P6J power supply. Do not
open, expose, modify or touch internal circuitry of the power supply.