Page 61
§ 7.
Trouble Shooting
When trouble occurs, a customer is required to record error message with "Error Code." That code
includes information to solve the trouble.
§ 7.1.
ERROR 01 Too Long Fiber
01-01
Left fiber is too long
01-02
Right fiber is too long
01-03
Both left and right fibers are too long.
Reason1:
Improper loading
Reason2
: Micro-dust
§ 7.2.
ERROR 02 Too Dark
02-01
X image brightness is not enough ( not zero )
02-02
X image brightness is not enough ( not zero )
Reason1:
Improper loading
Reason2:
Micro-dust
Reason2:
Mechanical adjustment (mirror or kyo utou).
Reason3:
LED or wiring damaged.
02-11
X image brightness is zero
02-12
Y image brightness is zero
Reason1:
CCD camera is damaged
Reason2:
CPU Board is damaged
Reason3:
LED connector is out of place
Reason4:
LED wire is damaged
Reason5:
LED is damaged
§ 7.3.
ERROR 03 Fiber too Dusty
03-01
Image processor cannot find fiber on the X image
03-02
Image processor cannot find fiber on the Y image
Reason1:
Micro-dust
Reason2:
cleaning arc weak (especially true for carbon coated fiber)
Reason3:
Wrong fiber mode
Содержание FSM-40S