Page 39
§ 4.10.2.
Confirmation of Swing-arm Travel
1.
Prepare and load two 250µm fibers into splicer.
2.
Refer to Fig. 4.10.3. Execute “Camera Position” command in [Secret Menu] and
push “SET” key
.
Note: X and Y motors automatically move to there outer limits at which time they are measured.
3.
Record the “U” and “D” values in both axes.
4.
Specification for “X” or “Y” movable range: both “U” and “D” values > 30 µm.
Camera Position
X
Y
182.7
278.5
374.2
94.6
190.2
285.9
97.2
193.0
8
299.2
94.1
190.0
285.8
Camera Position
X
Y
182.7
278.5
219.3
257.5
277.9
300.1
177.0
198.8
219.4
258.0
278.4
300.6
U: 38.5
D: 43.8
U: 42.8
D: 45.2
U: Top limit distance [µm]
D: Bottom limit distance [µm]
Fig. 4.10.3