Page 50
3.
Wait 1 minute.
4.
Refer to Fig. 4.18.6. Adjust A/D bit value up or down while checking temperature sensor output.
Adjustment should be complete in less than 30 seconds.
Specification = maximum heat
temperature 248 ~ 250ºC
.
5.
Press “ENT” key.
Heat Time
Heat
Temp erature
Heat temp. adjustment must be done in this period.
60 seconds
30 sec
.
Max. temperature is between 248 and 250ºC.
Adjusting point for cool finish temperature
must be between 100 and 105ºC.
Note: It is only one chance to check it. When the
temperature is out of specification, a re-adjustment
for AD100 is required. And, execute the program
from the first step.
Fig. 4.18.6
6.
Refer to Fig. 4.18.7. Adjust “100ºC” A/D bit value up or down while checking temperature sensor
output.
Specification = temperature 100 ~ 105ºC
. Note – Reading for adjustment should be taken
during the first decline from the high temperature point (refer to Fig. 4.18.6).
Note – A/D must be > 50 bit after adjustment. If not, refer to section 7.13 “Heater Trouble” to repair.
Adjust Temparature 250 deg
A/D 250 deg = 80 bit
Wait 59 sec.
Adjust Temparature 100 deg
A/D 100 deg = 220 bit
Fig. 4.18.7
Содержание FSM-40S