Page 52
§ 4.19.2.
Stuff Check
1.
Prepare and load two 250µm SM fibers into splicer.
2.
Refer to section 2.4.1. Execute “Stuff Check” in the [Secret Menu]
3.
The result of stuff check is displayed on the monitor.
Specification: 30±5 µm.
4.
If this value is no good, adjust the stuff speed by pushing “Up” or “Down” cursor key.
5.
Repeat steps 1 ~ 3.
§ 4.19.3.
Arc Calibration
1.
Refer to the instruction manual.
2.
Execute “Arc Calibration” mode in the main menu.
3.
If “Test again” appears on the monitor after executing arc calibration mode, repeat until “Test
again” message does not appear.
4.
If “Arc Calibration” is attempted 10 times with poor results. Execute “Replace Electrode” mode
without changing the electrodes.
5.
Repeat “Arc Calibration”.