Page 43
§ 4.14.
Install the Discharge Circuit
1.
Connect discharge circuit wiring to PCB connector CN24.
2.
Refer to section 3.2. Insert the discharge wires (anode and cathode) into the space between the
Mechanical base and the Base frame.
3.
Refer to Fig. 4.14.1. Insert the discharge wires into the electrode holders and tighten set screw
(M2×2, T=1.5).
4.
Attach discharge circuit to left side of Frame-Base with cap screw (M3×6, T=5, WS3, SLG).
Insert the discharge wire into the holes
Set screw M2×2, T=1.5
Fig. 4.14.1
Содержание FSM-40S
Страница 53: ...Page 53 5 Electrical 5 1 PCB Connection ...