Page 34
§ 4.9.3.
Image Alignment
1.
Refer to Fig. 4.9.5.
2.
Install special tool (26mm electrode) into rear (anode side). Note - The tip should not to touch the
fiber.
3.
Refer to section 3.1.1,. Install special tool (cap screw and a spring No.1053) into center of Focus
Adjuster. Gently rotate screw clockwise until it touches Kyoutou, then rotate the screw
counterclockwise ½ turn.
4.
Prepare and load two 250µm fibers into splicer.
5.
Refer to section 2.4.1 (page 1 of 4), execute “Camera Position” command in [Secret Menu].
Fig. 4.9.5
6.
Refer to Fig. 4.9.6.
7.
Loosen vertical position Cap screws (M3×10, WS3).
8.
Install 1.5mm block gauge between Kyoutou unit and Focus Adjuster.
9.
Apply pressure by pressing Kyoutou unit onto Focus Adjuster. Tighten Cap screw (M3×10, T =
temporary) “1” first and cap screw (M3×10, T = temporary) “2” second.
10.
Check X-axis and Y-axis positions.
Specification X=280 ± 10 line, Y=200 ± 10 line. Note –
Target X=285, Y=195.
11.
If out of specification, adjust Block Gauge thickness. Repeat steps 9 ~ 10 until the specification is
achieved.
12.
Tighten Cap screw (M3×10, T=12) “1” first and Cap screw (M3×10, T=12) “2” second.
13.
Refer to Fig. 4.9.7 for steps 14 ~ 21.
14.
Switch fiber view to “X large” view by pressing [X/Y] key.
15.
Enable focus “X” motor. Adjust the white band to one-third of fiber diameter.
16.
Switch fiber view to “Y large” view by pressing [X/Y] key.
17.
Enable focus “Y” motor. Adjust the white band to one-third of fiber diameter.
18.
Loosen horizontal position Cap screws (M3×10, WS3)
19.
Electrode cursor should be set at
line 376
.
20.
Adjust right side fiber end-face in the “Y” view by moving the Focus-Adjuster. Tighten cap screw
(M3×10, T=12) “1” first and Cap screw (M3×10, T=12) “2” second. Specification Y=376 ± 10 line.
Note – Target=366-376 line