![FujiFilm CR-IR 355RU Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/fujifilm/cr-ir-355ru/cr-ir-355ru_operation-manual_2343348059.webp)
A-3
Specifications
A
FCR XC-1 Operation Manual
First Edtion 11.2005 897N0565
A.1.2 Operating Performance
(1) Processing capacity
IP size
Processing type
14
"
×17
"
14
"
×14
"
10
"
×12
"
8
"
×10
"
24×30cm
18×24cm
Throughput
(Unit : Min. number of
processable IPs/hr.)
35
40
45
56
45
60
Feed/load time
(Unit : Max. time in
seconds)
101
88
80
64
79
60
• If the maximum dose applied on the IP exceeds 40mR (1 [mR]
≈
8.7 [µGy]), the cycle time increases as the X-ray
dose increases.
• The time required for changing the cassette is assumed 0 seconds.
• The erasure time is based on 25mR of X-ray dose (1 [mR]
≈
8.7 [µGy]) applied at exposures.
• The ST-VI type IP is used (the erasure time is extended for the ST-VN type IP).
(2) Time required for monitor image display
Time required for an image to be displayed on the monitor after the cassette has been inserted.
(Unit : Max. time in seconds)
Time required for
monitor image
display
(for each image
reading size)
Monitor
14
"
×17
"
14
"
×14
"
10
"
×12
"
8
"
×10
"
24×30cm
18×24cm
CR Console
52
46
41
33
41
31
(3) Time required for generating film prints
(Unit : Max. time in seconds)
Printer
(printout on a
14
"
×17
"
film)
Console
DRYPIX 7000
DRYPIX 4000
14"×17" IP
14"×14" IP
14"×17" IP
14"×14" IP
CR Console
135
130
155
145
* For one-image printout.
(4) Required specifications and other
Basic requirement for the CR
Noise (DQE)
: Should be within the range specified below.
Between 19.8% and 11.6% at 1cy/mm (258nC/kg of X-rays)
* The guaranteed initial center value is 17.0% for the six years of the equipment’s
effective life time.
Sharpness (MTF) : Should be within the range specified below.
Between 66.2% and 54.2% at 1cy/mm (825.6nC/kg of X-rays)
* The guaranteed initial center value is 60.0% for the six years of the equipment’s
effective life time.
Artifact
: Should be free of image unevenness that may affect the diagnosis.
Содержание CR-IR 355RU
Страница 2: ...ii FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 4: ...iv FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 40: ...4 4 Utility 4 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 50: ...5 10 Troubleshooting 5 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 56: ...6 6 Care and Maintenance 6 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 62: ...A 6 Speci cations A FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 66: ...B 4 IP Handling B FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 67: ...C 1 C FCR XC 1 Operation Manual First Edtion 11 2005 897N0565 Appendix C About Text Information Printed on Film...
Страница 72: ...C 6 C About Text Information Printed on Film FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 74: ...Y 2 Image Processing Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 107: ...YIV 1 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565 Appendix YIV Flexible Noise Control Software Parameters...
Страница 111: ...YIV 5 Flexible Noise Control Software Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 114: ...YIV 8 Flexible Noise Control Software Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 122: ...2 Maintenance and Inspection FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 124: ...2 Index FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 125: ...FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 126: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...