5-7
Troubleshooting
5
FCR XC-1 Operation Manual
First Edtion 11.2005 897N0565
Error code
Message title
Description
Corrective action
11530
Resend image
Since image could not be
transmitted, a resend is
attempted.
Check the conditions of the
console, connections cables, and
hub.
11531
Image send failure
Image could not be sent.
Press “Eject” and remove the
cassette.
11700
Dead erase lamp
Since some erasing lamp(s) is
dead, image erasure time is
extended. The erasing lamps
need inspection as soon as
possible.
Contact your local authorized
distributor.
11701
Erase incomplete
Since erasing lamps are dead,
image erasure is not possible.
Erasing lamps need replacing.
Contact your local authorized
distributor.
11702
Replace erase lamp
Erasing lamps need replacing.
Contact your local authorized
distributor.
11703
Life end. Long
erasure
Due to the end-of-life of erasing
lamps, the image erasure time is
extended. Lamp changing should
be done as soon as possible.
Contact your local authorized
distributor.
11704
Life end. Poor
erasure
Because of the end-of-life of
erasing lamps, image cannot be
erased. The erasing lamps need
replacing.
Contact your local authorized
distributor.
11710
Erase-lamp temp.
sensor fail
Since temperature sensor is
faulty, image cannot be erased.
The temperature sensor needs
replacing.
Contact your local authorized
distributor.
11720
Failed cooling fan
Since cooling fan for erasing
lamps is faulty, image cannot be
erased. The cooling fan needs
replacing.
Contact your local authorized
distributor.
11760
IP erase incomplete
Image on IP is not completely
erased.
Press “Eject.” Remove the
cassette once, insert it again, and
then erase the image.
11901
Worn air filter
Air filter is worn and needs
replacing as soon as possible.
Contact your local authorized
distributor.
11902
Worn brush roller
Brush roller is worn and needs
replacing as soon as possible.
Contact your local authorized
distributor.
11903
Worn damper
Damper is worn and needs
replacing as soon as possible.
Contact your local authorized
distributor.
11904
Worn erasure lamp &
filter
Lamp filter is worn and needs
replacing as soon as possible.
Contact your local authorized
distributor.
11905
Worn pump
Pump is worn and needs
replacing as soon as possible.
Contact your local authorized
distributor.
11907
Maintenance data
saving failure
Information for maintenance
checkup purposes cannot be
collected.
Start up the console that is
connected to the FCR XC-1.
Содержание CR-IR 355RU
Страница 2: ...ii FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 4: ...iv FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 40: ...4 4 Utility 4 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 50: ...5 10 Troubleshooting 5 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 56: ...6 6 Care and Maintenance 6 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 62: ...A 6 Speci cations A FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 66: ...B 4 IP Handling B FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 67: ...C 1 C FCR XC 1 Operation Manual First Edtion 11 2005 897N0565 Appendix C About Text Information Printed on Film...
Страница 72: ...C 6 C About Text Information Printed on Film FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 74: ...Y 2 Image Processing Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 107: ...YIV 1 FCR XC 1 Operation Manual First Edtion 11 2005 897N0565 Appendix YIV Flexible Noise Control Software Parameters...
Страница 111: ...YIV 5 Flexible Noise Control Software Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 114: ...YIV 8 Flexible Noise Control Software Parameters FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 122: ...2 Maintenance and Inspection FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 124: ...2 Index FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 125: ...FCR XC 1 Operation Manual First Edtion 11 2005 897N0565...
Страница 126: ...FUJIFILM MEDICAL SYSTEMS U S A INC 419 WEST AVENUE STAMFORD CT 06902 U S A...