Zero drift:
±1% of full scale per week
±2% of full scale per week (for ranges between 0–50
ppm and 0–200 ppm)
±2% of full scale per day (for ranges below 0–50 ppm)
Span drift:
±2% of full scale per week
±2% of full scale per day (for ranges below 0–50 ppm)
Response time for 90% FS response:
15 seconds electrical response
5HVSRQVHWLPHZKHQWKHJDVÀRZUDWHLV/PLQLV
within 60 seconds including replacement time of sample
gas.
*DV UHSODFHPHQW WLPH GHSHQGV RQ WKH QXPEHU RI
measuring components and range.
Standard Requirements for Sample Gas
Flow rate:
0.5L / min ±0.2L / min
Temperature:
WR&
Pressure:
10 kPa or less (Gas outlet should be opened to the
atmospheric air.)
Dust:
ȝJ1P
3
RUOHVVLQSDUWLFOHVL]HRIȝPRUOHVV
Mist:
Unallowable
Moisture:
%HORZDOHYHOZKHUHVDWXUDWLRQRFFXUVDW&FRQGHQVD
tion unallowable).
Corrosive component:
1 ppm or less
Standard gas for calibration:
Zero gas: Dry N
2
Span gas: a component same as the measuring target,
having concentration of 90–100% of its mea-
suring range (recommended). Gas beyond a
concentration of 100% FS is unusable.
In the case where an external zirconia O
2
analyzer is
installed and calibration is carried out on the same cali-
bration gas line as the other components:
Zero gas;
'U\DLURUDWPRVSKHULFDLU$WPRVSKHULFDLULV
not allowed if your measuring target includes
CO
2
.)
Span gas; For other than O
2
measurement, use a gas
that has concentration of 90–100% of its
measuring range. For O
2
measurement, use
1–2 vol% O
2
.
Installation Requirements
,QGRRUXVH6HOHFWDSODFHZKHUHWKHDQDO\]HUGRHVQRW
receive direct sunshine, wind and rain, or radiation from
hot substances. If such a place cannot be found, a roof
or cover should be prepared for protection.
$YRLGDSODFHZKHUHWKHDQDO\]HUUHFHLYHVKHDY\YLEUDWLRQ
6HOHFWDSODFHZKHUHDWPRVSKHULFDLULVFOHDQ
'LVFKDUJHWKHH[KDXVWJDVWRWKHVDIHSODFHWRWKHDWPR
sphere.
'RQRWXVHWKHDQDO\]HULQKD]DUGRXVDUHD
$OWLWXGHXSWR\DUGVP
EU Directive Compliance
LVD (2014/35/EU)
EN
61010-1
EN
62311
EMC (2014/30/EU)
EN 61326-1 (Table 2)
(1*URXS&ODVV$
(1&ODVV$
EN
61000-3-3
EN
61326-2-3
* The analyzer conforms to the EMC requirements only
when installed in a steel cabinet.
RoHS (2011/65/EU)
EN
50581