Password Set : Set the password used to move
from the parameter setting screen
to the maintenance mode.
Arbitrary 4-digit number can be
selected.
O
2
ref. Value
: Set the oxygen concentration
reference value at the time of
oxygen conversion calculation.
Settable in the range from 00 to
19%.
Limit
: Set the oxygen concentration limit
at the time of oxygen conversion
calculation. Settable in the range
from 01 to 20%.
* Refer to the O
2
conversion concentration value
in “5.3 Outline of display screen” for oxygen
conversion calculation procedure.
Station No.
: Set the station No. for MODBUS
communication. Settable
in the range from 00 to 31.
Range setting : Set/change the measurement range.
Description of each setting screen
Press
the or the
key to move the
cursor to the item whose setting is to be
changed.
The values for password, oxygen
conversion, limit, and station No. are
highlighted.
Press
the or the
key to change
the value to desired one, and then press
the
ENT
key.
Pay attention not to forget the password.
Otherwise you cannot enter the maintenance
mode.
CAUTION