3.2.2 Mounting input/output terminal module
0RXQWWKHLQSXWRXWSXWWHUPLQDOPRGXOHRQWKHSDQHOREVHUYLQJWKHIROORZLQJPHWKRG
(Note) To avoid the effect of noise generated from external units, mount the I/O terminal
module mounting plate on the panel for continuity at the mounting surface and con-
nect the panel to the same ground as the analyzer main unit.
Note) How to ground analyzer main unit and I/O terminal module
To avoid the effect of noises, etc. from external units, it is
recommended to ground them by the procedure described below.
Mount the analyzer and I/O terminal module on the
same panel.
Bring I/O terminal module sheet metal and
panel into continuity at ///// portion.
Bring analyzer main unit and panel into continuity
at ///// portion or rail mounting portion.
Ground the panel casing.
* Don’t separate the analyzer and I/O terminal module, and be sure to
ground them together.