6 - 33
Reset input
So long as close, resetting lasts.
At the edge of changing from closing to opening, the average action restarts.
Close (hold at least 1.5 sec.)
Open
Average value reset
This mode is used to clear O
2
average values and O
2
conversion average values and restarts aver-
aging. All average values are reset at a time. The indication value and output value is 0 ppm,
YRORUVRDWWKHWLPHRIWKHUHVHWLQSXW5HIHUWRWKHDYHUDJHSHULSG
Average period
It allows you to set an average period of the
average value of O
2
conversion
and O
2
average.
,WHQDEOHV\RXWRVHWDQDYHUDJHWLPHRIWR
minutes (1-minute step) or 1 to 4 hours (1-hour
step).
&KDQJLQJWKHVHWWLQJUHVHWVWKHDYHUDJHYDOXH
of O
2
conversion and O
2
average value. (Press-
ing the
ENT
validates the resetting only for
components whose setting was changed.)
Response time
The response time of the electrical system can
be changed.
Setting is available by components.
Note) It does not provide exact seconds for
the setting time, but it gives a guide
of the setting time.
7KHVHWWLQJYDOXHFDQEHPRGL¿HGDV
requested by the customer.