![Fuji Electric ZKJ Series Скачать руководство пользователя страница 48](http://html1.mh-extra.com/html/fuji-electric/zkj-series/zkj-series_instruction-manual_2341903048.webp)
6 - 8
To perform calibration for “both,” set the same calibration gas concentration for both
ranges.
CAUTION
When setting NOx and CO to “both”
Two cursors will appear in both ranges (Ch1 and Ch4).
Manual Calibration screen
(
)
ENT
End of Setting of calibration range
To close “Setting of Calibration Range” or
to cancel this mode midway, press the key.
A previous screen will return.
To close “Setting of Calibration Range”
ESC
Ch1
NOx
Ch2
SO
2
Range 1: 0 to 100 ppm
Range 2: 0 to 2000 ppm
Range 1: 0 to 100 ppm
Range 2: 0 to 2000 ppm
both
current
Ch1: Range 1 and Range 2 are calibrated together with zero and span calibration.
Ch2: Only currently displayed range is calibrated with zero and span calibration.
Example
2QWKH³FDOLEUDWLRQUDQJHVHOHFWLRQ´
VFUHHQWKDWDSSHDUVVHOHFW³ERWK´RU³FXU
-
UHQW´E\SUHVVLQJWKH
or the
key.
,I³ERWK´LVVHOHFWHG]HURRUVSDQFDOL
-
bration is performed with Range 1 and
5DQJHRIWKHVHOHFWHG&KLQWHUORFNHG
,I³FXUUHQW´LVVHOHFWHG]HURRUVSDQFDOL
-
bration is performed only for the range
GLVSOD\HGZKHQFDOLEUDWLRQRIVHOHFWHG&K
is performed.
Press
the
ENT
key after the selection, and
WKHVSHFL¿HGFDOLEUDWLRQLVSHUIRUPHG