9 - 2
Remote output holding:
$SSO\LQJWKHVSHFL¿HGYROWDJHRQWKHGHGLFDWHGWHUPLQDO
allows you to hold the output signal to the last value or
WKHXVHUVSHFL¿ROGLQJLVHIIHFWLYHZKLOHWKH
YROWDJHLVDSSOLHG9DOXHVLQGLFDWHGRQ/&'DUHQRWKHOG
Range changeover:
You can change between ranges by manually, automati-
cally, or remotely.
Manual: by key operation
$XWR
:KHQWKHPHDVXUHGYDOXHUHDFKHVDERYH
FS of the 1st range, the range automatically
VZLWFKHVWRWKHQGUDQJH:KHQWKHPHDVXUHG
value goes down below 80% FS of the 1st
range, the range automatically switches from
the 2nd range to the 1st range.
5HPRWH E\ WKH QRYROWDJH FRQWDFW LQSXW :KHQ WKH
remote range changeover input terminal dedi-
cated for each component is closed, the 1st
UDQJHLVHIIHFWLYH:KHQWKHWHUPLQDOLVRSHQHG
the 2nd range becomes effective.
5DQJHLGHQWL¿FDWLRQVLJQDO
<RXFDQFKHFNZKLFKUDQJHLVLQXVH:KHQWKHVWUDQJH
LV XVHG WKH UDQJH LGHQWL¿FDWLRQ VLJQDO RXWSXW WHUPLQDO
DOORFDWHGIRUHDFKFRPSRQHQWLVFORVHG:KHQWKHQG
range is used, the terminal is opened.
Auto calibration:
This function requires standard gas cylinders for zero
and span calibration and solenoid valves for opening/
FORVLQJWKHJDVÀRZOLQH:KHQWKLVIXQFWLRQLVDFWLYDWHG
the analyzer opens and closes the solenoid valve driving
contact periodically at preset cycle.
Auto calibration cycle setting:
1 hour to 99 hours (in increments of 1 hour) or 1 day to
40 days (in increments of 1 day).
*DVÀRZWLPHVHWWLQJ
The time during which calibration gas is supplied.
60 seconds to 900 seconds (in increments of 1 second)
Auto calibration remote start:
You can start one-time auto calibration if you open
the auto calibration remote start input terminal for 1.5
seconds or longer and then close it. Calibration gas is
GUDZQIRUWKHWLPHVHWLQWKH³JDVÀRZWLPHVHWWLQJ´IRU
$XWRFDOLEUDWLRQVHHWKHSUHYLRXVLWHP
Auto zero calibration:
This function requires a standard gas cylinder for zero
calibration and a solenoid valve for opening/closing
WKH JDV ÀRZ OLQH :KHQ WKLV IXQFWLRQ LV DFWLYDWHG WKH
analyzer opens and closes the solenoid valve driving
contact periodically at preset cycle. The cycle for the
auto zero calibration and that for the auto calibration
can be different.
$XWR]HURFDOLEUDWLRQF\FOHVHWWLQJ
1 hour to 99 hours (in increments of 1 hour) or 1 day
to 40 days (in increments of 1 day).
*DVÀRZWLPHVHWWLQJ
The time during which calibration gas is supplied
60 seconds to 900 seconds (in increments of 1 second)
High/low limit alarm:
:KHQDPHDVXUHGYDOXHKDVJRQHEH\RQGWKHXSSHUOLPLW
or below the lower limit, the analyzer closes the contact
to emit an alarm signal.
Instrument error contact output:
The contact is closed when an analyzer error (error No.
1 or 10) occurs.
Calibration error contact output:
The contact is closed if a calibration error (error No. 4,
5, 6, 7, or 9) occurs.
Auto calibration status contact output:
The contact is closed during auto calibration.
Pump ON/OFF contact output:
The contact is closed during measurement, and opened
GXULQJFDOLEUDWLRQVRWKDWWKHVDPSOHJDVÀRZLVVWRSSHG
during calibration
Optional Functions
O
2
conversion:
Conversion of measured NO, CO, and SO
2
gas concen-
trations into values at reference O
2
concentration
&RQYHUVLRQIRUPXOD& î&V
C: Sample gas concentration after O
2
conversion
Cs: Measured concentration of sample gas
Os: Measured O
2
FRQFHQWUDWLRQOLPLWFRQ¿JXUDEOHZLWKLQ
1–20%)
On: Reference O
2
FRQFHQWUDWLRQ FRQ¿JXUDEOH ZLWKLQ
0–19%)
Average value after O
2
conversion and O
2
average valu
ecalculation:
The analyzer can take measurement every 30 seconds,
and calculate the moving average of:
a) the instantaneous concentrations after O
2
correc-
tion or
b) the instantaneous O
2
concentrations.
You can set the period for averaging in the range of 1–59
min (in one-minute increment) or 1–4 hour (in one-hour
increment).
The analyzer transmits the moving average output every
30 seconds.
Average value resetting:
You can reset the average values by short-circuiting the
average value resetting input terminal for 1.5 seconds
or longer. You can start averaging over by opening it.
CO concentration peak count alarm:
(available option only for CO and O
2
analyzer)
The analyzer counts the number of times that the CO
instantaneous value has reached the concentration limit
\RXVHW:KHQWKHFRXQWSHUKRXUKDVUHDFKHGWKHOLPLW
you set, the alarm is triggered.
Communication function:
RS-232C
9-pin D-sub connector
Half-duplex bit serial, Start-stop synchronization
Modbus RTU™ protocol
The communication function allows the analyzer to read
and write parameters, and read measured concentration
values and instrument status.
For connection to RS-485 interface, an RS232C–RS485
converter is required.
Performance
Repeatability:
±0.5% of full scale
±1% of full scale (for ranges below 0–50 ppm)
Linearity:
±1% of full scale
21–On
21–Os