7.3 Zero calibration
Refer to “6.1 Zero calibration”.
7.4 Span calibration
Refer to “6.2 Span calibration”.
7.5 Replacement of packing
Replace the packing, referring to the following figure.
Replace the receiver packing and silicon packing C every two years.
7.6 Light axis adjustment
Refer to “4.7 Angle adjustment” and “4.11 Adjustment of light quantity”.
7.7 When the device is not used for a long period
If the device is attached to the stack without performing air-purge for a long period, parts such as the receiver
and transmitter units might be damaged. When you do not use if for a long period, detach the device from the
stack and keep it in a place not exposed to high temperature and risk of corrosion.