INZ-TN1ZSS-E
- 43 -
Explanation of alarm range
Range1 H-Limit : Alarm output is provided only when a value exceeds the Range1 H-Limit.
Range1 L-Limit : Alarm output is provided only when a value is lower than Range1 L-Limit.
Range1 H/L-Limit : Alarm output is provided when a value exceeds the Range1 H-Limit or it is lower
than Range1 L-Limit.
Follow the same procedure for Range2.
6.3.4 Hysteresis setting
Set the hysteresis to prevent possible chattering of the alarm output near the alarm setting value.
Point the to “Hysteresis” by the
key or
the
key, and press the
key.
(
)
Change the numeric value by the
key or
the
key, and move the digit by the
key.
Press the
ENT
key to make the hysteresis valid.
(
)
ENT
Alarm Setting
H
-
Limit
ON
Range1
15
ppm
Range2
ppm
L-Limit
ppm
ppm
Hysteresis
%FS
Analog Output
Range1 H-Limit
Alarm Record
50.00
30.00
10.00
00.00
Range1
Range2
Alarm Setting
H
-
Limit
OFF
Range1
15
ppm
Range2
ppm
L-Limit
ppm
ppm
Hysteresis
%FS
Analog Output
Range1 H-Limit
Alarm Record
50.00
30.00
10.00
00.00
Range1
Range2
Alarm Setting
H
-
Limit
ON
Range1
1
5
ppm
Range2
ppm
L-Limit
ppm
ppm
Hysteresis
%FS
Analog Output
Range2 H-Limit
Alarm Record
50.00
30.00
10.00
00.00
Range1
Range2