6. ZERO CALIBRATION AND SETTING
6.1 Zero calibration
Select “Zero Calibration” from the “Menu” screen and press the
ENT
key.
AO Select
2006-04-01
12:00:00
Menu
Zero Calibration
Alarm Log
Display
/
Output
Password
*****
Parameter Setting
Span Calibration
Alarm Setting
Input Signal
Output Hold
ENT
6.1.1 Preparation
CAUTION
•
The laser beam is the invisible infrared light. Do not watch the laser
beam directly or scattering light.
•
Do not watch the laser beam directly with the optical measuring
device. Otherwise, it may cause serious damage to your eyes.
Parts name
Quantity
Remarks
Calibration gas cell
1
To be ordered separately
Receiver/transmitter cable (for cali-
bration)
1
To be ordered separately
Receiver/transmitter cable
1
To be ordered separately
Zero gas (N
2
)
1
To be ordered separately
Pressure adjuster (regulator)
1
To be ordered separately
Pipe (Teflon tube, etc)
Several m
φ
10 × 8 or more
Joint (Rc1/4)
2
For calibration gas cell
Flow meter
1
2L/min or more
Thermometer
1
4 to 20mA output, for temperature correc-
tion
Pressure gauge
1
4 to 20mA output, for pressure correction
Others (joint, tools, etc)
(1) Turn OFF the power.
(2) Remove the receiver box and the transmitter box using the hexagonal wrench. Perform removing when
the provided equipment is stopped. Never remove boxes while the equipment is operated.
Otherwise, hot temperature gas may blow out.
2007-4-1
12:00:00
Zero Calibration
ZeroCal.Start
Cell Length
LastCal.Day
2006/09/01
Cal.Finish
Meas.
00.12
1000
mm
ppm
Cal.Gas Value
0000
ppm
Component
HC
ℓ