INZ-TN1ZSS-E
- 51 -
Remote average value reset (DI1 terminal)
Apply rectangular waveform voltage (pulse width: 2.0 sec or more, pulse interval: 2.0 sec or more) to the
input terminal of remote average value reset (DI1 terminal, option) to reset the average value.
Reset is kept during short-circuit.
When switching the action from short-circuiting to opening, average value action starts.
6.5.4 Key lock
When key lock is ON, any key operation except the key lock OFF cannot be set.
2006-04-01
12:00:00
Menu
Zero Calibration
Alarm Log
Display
/
Output
Password
*****
Parameter Setting
Span Calibration
Alarm Setting
Output Hold
AO Select
Input Signal
When the key lock is turned ON, any screens
except “Parameter Setting” cannot be dis-
played, even if the item is selected and the
ENT
key is pressed on the “Menu” screen.
Any items except “Key Lock” cannot be vali-
dated, even if the item is selected and the
key is pressed on the “Parameter Setting” screen.
2.0s or more
Remote hold signal
Open
Short-circuit
Starts average value reset
Parameter Setting
Date/Time
07
/
01
/
01
00:00:00
Average Period
01
Min.
Reset Avg.Gas
Disp.Off
ON
09
Min.
Path lengths
01000
mm
Key Lock
OFF
HC
ℓ
2007-4-1
12:00:00