6.5.1 Setting of stack lengths
Enter the lengths of the stack (inner diameter) where the receiver unit and the transmitter unit are attached. It
does not include the lengths of companion flange. This value has a direct effect on the measurement value, so
be sure to enter correctly. Otherwise, measurement value will not be properly displayed or output. Enter the
value in the millimeter. (Input range is from 500 to 15000mm.)
Explanation of concentration output
The analyzer is calculated and output as a standard of the 1m of stack lengths (optical path lengths).
Based on Lambert-Beer Law, absorption concentration varies depending on the stack lengths, so be sure to
enter the stack lengths correctly.
Point the to “Path Lengths” by the
key
and the
key, and press the
key.
Parameter Setting
Date/Time
07
/
01
/
01
00:00:00
Average Period
01
Min.
Reset Avg.Gas
Disp.Off
ON
09
Min.
Path lengths
01000
mm
Key Lock
OFF
HC
ℓ
2007-
4
-1
12:00:00
Parameter Setting
Date/Time
07
/
01
/
01
00:00:00
Average Period
01
Min.
Reset Avg.Gas
Disp.Off
ON
09
Min.
Path lengths
01000
mm
Key Lock
OFF
HC
ℓ
2007-4-1
12:00:00