5.3.1.9 Alarm types
Following types of alarm are displayed.
Alarm display
Alarm contents
Probable causes
LD Failure
Laser is faulty
•
Failure caused by laser durability
LD Temp. Error
Peltier which is cooling the laser
cannot control the set temperature.
•
Peltier failure
•
Thermocouple failure
•
The receiver unit and the transmitter
unit are used at an installation location
exceeding the set range.
•
The unit is used in an environment
where gas temperature is beyond the
specification.
Low Light Trans.
Light transmission required for the
measurement cannot be obtained.
•
Light transmission is insufficient in an
environment with high dust.
•
Light transmission is insufficient in an
environment with high water vapor.
•
Contamination of window and conden-
sation are caused by insufficient air
purge.
•
Optical path is blocked due to dust.
•
Optical axis is deflected due to vibra-
tion.
•
Optical axis is deflected due to distor-
tion of a stack.
•
Optical axis is deflected due to external
faults.
Connection Error
Communication between receiver
unit and transmitter unit does not
occur properly.
•
Break of wiring
•
Wrong wiring of RS-432 terminal
•
Poor contacts of the connector unit
•
CPU board failure
•
PD digital board failure
High Gas Temp.
It is reported when exceeded gas
temperature is detected.
•
The actual gas temperature is more than
450°C.
•
The value set to analog input on the
“AO Select” screen is not correct.
Out of Range
It is reported when exceeded gas
pressure is detected.
•
Actual gas pressure is outside the speci-
fication range.
•
The value of the input range set on the
“AO Select” screen is not correct.
AI Under
Input channel on the “AO Select”
screen.
•
AI terminal is not connected to the ex-
ternal input device when channel setting
is set on the “AO Select” screen.
•
AI terminal and setting channel do not
match.
•
Input is not 4 to 20mA DC.
Box Temp. Warning It is reported when the temperature
in the receiver unit and the trans-
mitter unit exceeds the temperature
for normal operation.
•
The unit is used at an installation loca-
tion exceeding the set range.
•
The unit is used in an environment
where gas temperature is beyond the
specification.
•
Insufficient air purge causes rise in
temperature.
•
The distance between the Receiver /
Transmitter unit and the stack is not
maintained sufficiently.