Operating Manual EMGZ421
23
Offset channel B
Use:
This parameter stores the value determined with
Find offset B
in
[Digit]. It is not necessary to note this parameter, because a new offset
adjustment is done very easy; also when changing the whole
electronic unit.
The offset can also be inputted by using the
↑
↓
←
keys.
Range:
-4000
to
4000
Default:
0
Increment:
1
Unit:
[Digit]
Gain channel B
Use:
This parameter stores the value determined with
Calibration B
, resp.
you can input a value calculated using the formulas described under
„8.2 Calibrating the measuring amplifier“, if the
material tension
cannot be simulated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Gain 1 channel A
Use:
Identical
with
Gain channel A
, but the value stored here was
determined by
Calibration gain 1 -> A
. The value stored here is used
if the digital input „Gain switching A“ is activated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Gain 1 channel B
Use:
Identical
with
Gain channel B
, but the value stored here was
determined by
Calibration gain 1 -> B
. The value stored here is used
if the digital input „Gain switching B“ is activated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]
Nominal force sensor A
Use:
To get the correct force value, the electronic unit has to know the
nominal force of the sensors.
The parameters
Nominal force sensor A
and
Nominal force sensor B
have to be set to the same value.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N,kN]
Nominal force sensor B
Use:
Identical
with
Nominal force sensor A
.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N,kN]