Operating Manual EMGZ421
26
Scale output 1 (A–B)
Use:
At default setting of 0.50, the output 1 (±10V) provides the nominal
signal level (10V) when the difference A–B is reaching the nominal
force of the sensors. If the scale value stored here is reduced, the
nominal output signal is reduced too; if the scale value is enlarged, the
nominal output signal is enlarged too.
Range:
0.01
to
10.00
Default:
0.50
Increment:
0.01
Unit:
[-]
Scale output 2 (A+B)
Use:
At default setting of 0.50, the output 2 (0...10V) provides the nominal
signal level (10V) when the sum A+B is reaching the nominal force of
the sensors. If the scale value stored here is reduced, the nominal
output signal is reduced too; if the scale value is enlarged, the nominal
output signal is enlarged too.
Range:
0.01
to
10.00
Default:
0.50
Increment:
0.01
Unit:
[-]
Scale output 3 (A)
Use:
At default setting of 1.00, the output 3 (0...5V) provides the nominal
signal level (5V) when the feedback A is reaching the nominal force
of the sensors. If the scale value stored here is reduced, the nominal
output signal is reduced too; if the scale value is enlarged, the nominal
output signal is enlarged too.
Range:
0.01
to
10.00
Default:
1.00
Increment:
0.01
Unit:
[-]
Scale output 4 (B)
Use:
At default setting of 1.00, the output 4 (0...5V) provides the nominal
signal level (5V) when the feedback B is reaching the nominal force
of the sensors. If the scale value stored here is reduced, the nominal
output signal is reduced too; if the scale value is enlarged, the nominal
output signal is enlarged too.
Range:
0.01
to
10.00
Default:
1.00
Increment:
0.01
Unit:
[-]
Language
Use:
With this parameter, the language in the display can be chosen.
Range:
English, German, French, Italian
Configuration of key
Use:
This parameter defines wheter the free key (
↓
key) is assigned to the
Tare function or not. (Refer to „8.3 Additional settings“)
Range:
None, TareA, TareB, TareA+B
Default:
None