Operating Manual EMGZ421
22
Calibration B
Use:
In this parameter, using the
↑
↓
keys you can input the force value
referring to half of the calibration load you applied to the sensor (force
feedback value). The processor then calculates the actual gain value
and stores it under parameter
Gain channel B.
Notice: The input can be aborted with the
←
key. In this case the
previously saved value remains.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N,kN]
Calibration gain 1 -> A
Use:
Identical
with
Calibration A
, but the result is stored under parameter
Gain 1 channel A
. This gain value is used if the digital input „Gain
switching A“ is activated.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N,kN]
Calibration gain 1 -> B
Use:
Identical
with
Calibration B
, but the result is stored under parameter
Gain 1 channel B
. This gain value is used if the digital input „Gain
switching B“ is activated.
Range:
1
to
9999
Default:
1000
Increment:
1
Unit:
[N,kN]
Offset channel A
Use:
This parameter stores the value determined with
Find offset A
in
[Digit]. It is not necessary to note this parameter, because a new offset
adjustment is done very easy; also when changing the whole
electronic unit.
The offset can also be inputted by using the
↑
↓
←
keys.
Range:
-4000
to
4000
Default:
0
Increment:
1
Unit:
[Digit]
Gain channel A
Use:
This parameter stores the value determined with
Calibration A
, resp.
you can input a value calculated using the formulas described under
„8.2 Calibrating the measuring amplifier“, if the
material tension
cannot be simulated.
Range:
0.100
to
9.000
Default:
1.000
Increment:
0.001
Unit:
[-]