Operating Manual EMGZ421
28
11
Trouble shooting
Error
Cause
Corrective action
„Err1“ is displayed:
A/D-converter receives
Force sensor A is wrong
connected
Exchange wires on terminals z6 / z8 (resp. 2 /
3)
values < -9.7mV
continuously
Parting of the cable
Replace connection cable between force
sensor A and measuring amplifier
„Err2“ is displayed:
A/D-converter receives
Force sensor A is wrong
connected
Exchange excitation and signal (terminals z4
... z10 resp. 1 ... 4)
values > 9.7mV
continuously
Short circuit in the plug
or connection cable
Check and correct wiring to sensor A
Force sensor overload
Use sensor with higher nominal force
Force sensor has too
much sensitivity
Set parameter
sensitivity sensor A
to the
correct value or use other sensor
„Err3“ is displayed:
A/D-converter receives
Force sensor B is wrong
connected
Exchange wires on terminals d6 / d8 (resp. 7 /
8)
values < -9.7mV
continuously
Parting of the cable
Replace connection cable between force
sensor B and measuring amplifier
„Err4“ is displayed:
A/D-converter receives
Force sensor B is wrong
connected
Exchange excitation and signal (terminals d4
... d10 resp. 6...9)
values > 9.7mV
continuously
Short circuit in the plug
or connection cable
Check and correct wiring to sensor B
Force sensor overload
Use sensor with higher nominal force
Force sensor has too
much sensitivity
Set parameter
sensitivity sensor B
to the
correct value or use other sensor
„Err5“ is displayed:
Output 1 (A–B) shows
minimum continuoulsy
Offset badly adjusted
Proceed for offset adjustment again
„Err6“ is displayed:
Output scaling badly
adjusted
Set parameter
Scale output 1
to appropriate
value
Output 1 (A–B) shows
Offset badly adjusted
Proceed for offset adjustment again
maximum continuoulsy
Gain badly adjusted
Proceed for sensor calibration again
Feedback value is > 0 even
though material is loosely
Offset badly adjusted
Proceed for offset adjustment again
Feedback value isn’t stable
even though
Cut-off frequency of the
filters set too high
Adjust cut-off frequency (ref. to „8.3
Additional settings“)
material tension doesn’t
change
Ground terminal ogf the
output isn’t 0V
Connect Gnd terminal of the output (terminal
z22 resp. 14) with earth (terminal z24 resp.
PE)
Feedback value doesn’t
correspond with the
effective material tension
Gain badly adjusted
Proceed for sensor calibration again
No message on the display
Display contrast setting
is bad
Set display potentiometer correctly. (It is
located on the processor board on the upper
right edge beside the ribbon connector)
Fuse blown
Replace fuse on power supply
Power supply not
correct
Check / correct power supply
Electronic unit defect
Contact FMS customer service