Operating Manual EMGZ421
18
9.4
Write parameter
command answer
purpose
WP01XXXXX<CR>
PACC<CR> / FAIL<CR>
Offset channel A
WP02X.XXX<CR>
PACC<CR> / FAIL<CR>
Gain channel A
WP03XXXXX<CR>
PACC<CR> / FAIL<CR>
Offset channel B
WP04X.XXX<CR>
PACC<CR> / FAIL<CR>
Gain channel B
WP05X.XXX<CR>
PACC<CR> / FAIL<CR>
Gain1 channel A
WP06X.XXX<CR>
PACC<CR> / FAIL<CR>
Gain1 channel B
WP07XXXX<CR>
PACC<CR> / FAIL<CR>
Nominal force sensor A
WP08XXXX<CR>
PACC<CR> / FAIL<CR>
Nominal force sensor B
WP09X<CR>
PACC<CR> / FAIL<CR>
Unit of force
WP10X.X<CR>
PACC<CR> / FAIL<CR>
Sensitivity sensor A
WP11X.X<CR>
PACC<CR> / FAIL<CR>
Sensitivity sensor B
WP13XXXXX<CR>
PACC<CR> / FAIL<CR>
Min. limit value A+B
WP14XXXXX<CR>
PACC<CR> / FAIL<CR>
Max. limit value A+B
WP15XXXXX<CR>
PACC<CR> / FAIL<CR>
|A–B| > limit
WP18XX.X<CR>
PACC<CR> / FAIL<CR>
Lowpass display
WP19XXX.X<CR>
PACC<CR> / FAIL<CR>
Lowpass output 1 (A–B)
WP20XXX.X<CR>
PACC<CR> / FAIL<CR>
Lowpass output 2 (A+B)
WP21XXX.X<CR>
PACC<CR> / FAIL<CR>
Lowpass output 3 (A)
WP22XXX.X<CR>
PACC<CR> / FAIL<CR>
Lowpass output 4 (B)
WP23XX.XX<CR>
PACC<CR> / FAIL<CR>
Scale output 1 (A–B)
WP24XX.XX<CR>
PACC<CR> / FAIL<CR>
Scale output 2 (A+B)
WP25XX.XX<CR>
PACC<CR> / FAIL<CR>
Scale output 3 (A)
WP26XX.XX<CR>
PACC<CR> / FAIL<CR>
Scale output 4 (B)
WP28X<CR>
PACC<CR> / FAIL<CR>
Language
WP29X<CR>
PACC<CR> / FAIL<CR>
Config. of key
WP39XXX<CR>
PACC<CR> / FAIL<CR>
Identifier
WP40X<CR>
PACC<CR> / FAIL<CR>
Baud rate RS232
WP41X<CR>
PACC<CR> / FAIL<CR>
7 or 8 data bit
WP42X<CR>
PACC<CR> / FAIL<CR>
1 or 2 stop bit
WP43X<CR>
PACC<CR> / FAIL<CR>
Parity bit RS232
All parameter numbers and input ranges refer to the parameter list. Depending on the
value being ok or not, the electronic unit replys PACC<CR> (value accepted) or
FAIL<CR> (value not accepted).