190 Series III
Service Manual
52
Input D LF-HF Gain Calibration Points (see
) 4-channel models only
CL 0675
0.5 Vpp square wave, 1 kHz
SCOPE edge, 0.5 Vpp, 1 kHz
CL 0475
100 mVpp square wave, 1 kHz
SCOPE edge, 100 mVpp, 1 kHz
CL 0590
2.5 Vpp square wave, 1 kHz
SCOPE edge, 2.5 Vpp, 1 kHz
CL 0586
2.5 Vpp sine wave, 50 kHz
SCOPE levsine, 2.5 Vpp, 50 kHz
CL 0587
2.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 2.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0486
500 mVpp sine wave, 50 kHz
SCOPE levsine, 500 mVpp, 50 kHz
CL 0487
0.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 500 mVpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0466
100 mVpp sine wave, 50 kHz
SCOPE levsine, 100 mVpp, 50 kHz
CL 0467
100 mVpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 100 mVpp,
501 MHz
381 MHz
151 MHz
111 MHz
Input AB Position Calibration Points (see
) all models
CL 0637
500 mV, 1 kHz, MODE VOLT
Input AB Gain and Position Calibration Points (see
CL 0504
500 mV, 1 kHz, MODE VOLT
SCOPE volt, 500 mV, 1 kHz
CL 0624
Open inputs, Calibrator STBY
no signal
CL 0673
50 V, 1 kHz, MODE volt, (set 5502A to
OPR!)
SCOPE volt, 50 V, 1 kHz (set 5502A
to OPR!)
Input Pos ABCD (AB) Calibration Points
CL 0619
no signal
Input ABCD (AB) Noise F FBW Calibration Points
CL 0850
no signal
Table 14. Calibration Points (cont.)
Cal Step
DUT Input Signal
5502A Setting