190 Series III
Service Manual
44
Figure 16. 5502A SCOPE Output to Test Tool Input D
Input AB Position (All Models)
To do the Input AB Position calibration:
1. Press
2
to select calibration adjustment step
CL 0637
(Pos AB).
2. Set up the test shown in
with the 5502A set to supply 1 kHz square wave (SCOPE,
MODE volt, SCOPE Z 1 M
Ω
), of 500 mV to channel A and B.
3. Press
3
to start the calibration.
Wait until the display shows calibration status
CL 0637: READY
.
Continue at the
Input AB LF-HF Gain and Position
section.
Figure 17. Test Tool Input ABCD to 5502A Scope Output
A
D
C
B
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
PM9585 - 50
:
USE 50
:
TERMINATION
4 CHANNELS
For 190-502 and -504
use only TRM50!
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
CONNECT TO CHANNEL A, B, C, D IN PARALLEL
50
:
TERMINATION
NO
A
D
C
B
PM9093
PM9092
A
B
METER
P
M
9
0
9
3
PM9092
PM9091
2 CHAN. + METER
4 CHANNELS