ScopeMeter® Test Tool
Calibration Adjustment
43
Wait until the display shows calibration status
CL 0549: READY
.
5. Press
2
to select the next calibration step, set the 5502A to the next calibration
point signal, and start the calibration.
Continue through all calibration points of
Input C LF-HF Gain
.
6. Wait until the display shows calibration status
CL 0465: READY
.
7. When you are finished, set the 5502A to Standby.
8. Continue at the
section.
Input D LF-HF Gain
To do the Input D LF-HF Gain calibration:
1. Press
2
to select the first calibration step in
,
Input D LF-HF Gain
.
2. Connect Ch. D of the Test Tool to the 5502A. See
3. Set the 5502A SCOPE output to source the signal required for the first calibration point in
Input D LF-HF Gain
(CL 0675, Pos D Fast).
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
5. Press
3
to start the calibration.
Wait until the display shows calibration status
CL 0569: READY
.
6. Press
2
to select the next calibration step, set the 5502A to the next calibration
point signal, and start the calibration.
Continue through all calibration points of
Input D LF-HF Gain
.
7. Wait until the display shows calibration status
CL 0467: READY
.
8. When you are finished, set the 5502A to Standby.
9. Continue at the
Input AB Position (All Models)
section.