ScopeMeter® Test Tool
Performance Verification
13
h. Press
D
, then press
4
(INPUT D OPTIONS …).
i. Select
Attenuator: Normal | Bandwidth: 10 kHz
for the available setting in the
ScopeMeter (2 mV/d and 5 mV/d) or
| Bandwidth: 20 MHz
(other ranges).
j. Press
I
to clear the softkey menu and see the full 8-divisions display.
3. Press
P
.
4. Press
2
(– READING …) and select with
1
(– READINGS) and the
h i
:
Reading 1, on A, V ac
Reading 2, on B, V ac
Reading 3, on C, V ac
Reading 4, on D, V ac
5. Use
V
to change the time base and lock on 20
μ
s/div for the 20 kHz signal and on
10 ms/div for the 60 Hz signal.
6. Use
j
and
k
to select the manual vertical ranging. Set the input A and B sensitivity range
to the first test point in
The sensitivity ranges are indicated in the lower display edge.
7. Set the 5502A to source the appropriate ac voltage.
Observe readings
A
,
B
,
C
, and
D
and check to see if they are within the range shown under
the appropriate column.
8. Continue through the test points.
When you are finished, set the 5502A to 0 (zero) Volt and Standby.