ScopeMeter® Test Tool
Calibration Adjustment
37
Display Messages and Key Functions
When the Test Tool is in the calibration mode, only the
1
to
4
softkeys, the
O
key, and the
I
key are active, unless otherwise stated.
The calibration adjustment menu shows the actual calibration step (name and number) and its
status:
Cal Name (CL nnnn) %:Status (…)
Note
1
and
2
are disabled whenever they can harm the process.
Cal Name
Name of the selected calibration step, for example, WarmingUp
(CL nnnn)
Number of the calibration step
%
Progress %
Status (…) can be:
IDLE (valid)
After (re)entering this step, the calibration process is not started. The calibration
data of this step are valid. This means that the last time this step was done, the
calibration was successful. It does not necessarily mean that the unit meets the
specifications related to this step.
IDLE (invalid)
After (re)entering this step, the calibration process is not started. The calibration
data are invalid. This means that the last time this step was done, the calibration was
not successful. Most probably the unit will not meet the specifications if the actual
calibration data are saved.
BUSY aaa%
bbb%
Calibration adjustment step in progress; progress % for Input A and Input B. During
Warming-Up, the elapsed time is shown.
READY
Calibration adjustment step finished.
Error :xxxx
Calibration adjustment failed, due to wrong input signal(s) or because the Test Tool is
defective.
If the error code is <5000 you can repeat the failed step.
If the error code is
≥
5000 you must repeat the complete final calibration (start at
Warming-Up 2, Warm-Up Final, and ADC Timing
The functions of the keys are:
1
PREVIOUS
select the previous step (if applicable)
2
NEXT
select the next step (if applicable)
3
CALIBRATE
start the calibration adjustment of the actual step
4
EXIT
leave the calibration mode (turn off and turn on the
tool to make sure that the tool operation is normal)