ScopeMeter® Test Tool
Performance Verification
7
21. Press
1
SAVE... .
22. Use
hi
to select SETUP.
23. Press
E
.
24. Use
hi
to select OK SAVE.
Remember the name under which the settings are saved (for example, SCOPE 1).
25. Press
E
to save the settings.
26. Press
H
to leave the Hold mode.
Scope Input A, B, C, D Tests
Input A, B, C, D Vertical Accuracy Test
XW
Warning
To prevent possible electrical shock, fire, or personal injury, make sure that the
calibrator is in standby mode before making any connection between the
calibrator and the Test Tool. Dangerous voltages are present on the calibration
source and connection cables during these steps.
Note
The test steps for channels C and D are only for the 4-channel models.
To test:
1. Connect the Test Tool to the 5502A as shown in
. The vertical channels A, B, C, and
D are checked in succession and one waveform is on the display at a time to facilitate
amplitude adjustment.
Figure 2. Test Tool Inputs to 5502A Normal Output
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
A
D
C
B
PM9091
PM9081
CONNECT TO CHANNEL A, B, C, D IN SUCCESSION
50
:
TERMINATION
NO
A
B
METER
PM9091
2 CHAN. + METER
4 CHANNELS