ScopeMeter® Test Tool
Calibration Adjustment
51
CL 0483
0.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 0.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0462
100 mVpp sine wave, 50 kHz
SCOPE levsine, 100 mVpp, 50 kHz
CL 0463
100 mVpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 100 mVpp,
501 MHz
381 MHz
151 MHz
111 MHz
Input C LF-HF Gain Calibration Points (see
) 4-channel models only
CL 0656
0.5 Vpp square wave, 1 kHz
SCOPE edge, 0.5 Vpp, 1 kHz
CL 0455
100 mVpp square wave, 1 kHz
SCOPE edge, 100 mVpp, 1 kHz
CL 0550
2.5 Vpp square wave, 1 kHz
SCOPE edge, 2.5 Vpp, 1 kHz
CL 0584
2.5 Vpp sine wave, 50 kHz
SCOPE levsine, 2.5 Vpp, 50 kHz
CL 0585
2.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 2.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0484
500 mVpp sine wave, 50 kHz
SCOPE levsine, 500 mVpp, 50 kHz
CL 0485
0.5 Vpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 0.5 Vpp,
501 MHz
381 MHz
151 MHz
111 MHz
CL 0464
100 mVpp sine wave, 50 kHz
SCOPE levsine, 100 mVpp, 50 kHz
CL 0465
100 mVpp sine wave
Fluke 190-502/504: 501 MHz
Fluke 190-202/204/190M-2/4: 381 MHz
Fluke 190-102/104: 151 MHz
Fluke 190-062: 111 MHz
SCOPE levsine, 100 mVpp,
501 MHz
381 MHz
151 MHz
111 MHz
Table 14. Calibration Points (cont.)
Cal Step
DUT Input Signal
5502A Setting