190 Series III
Service Manual
42
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
5. Press
3
to start the calibration.
Wait until the display shows calibration status
CL 0529: READY
.
6. Press
2
to select the next calibration step, set the 5502A to the next calibration
point signal, and start the calibration.
Continue through all calibration points for
Input B LF-HF Gain and
wait until the display
shows calibration status
CL 463: READY
.
7. When you are finished, set the 5502A to Standby.
8. Continue at the
section.
Input C LF-HF Gain
Sections
Input C LF-HF Gain
and
Input D LF-HF Gain
are for 4-channel ScopeMeters (190-104,
190-204, 190M-4, 190-504, and MDA-550-III models). For 2-channel models, see
.
To do the Input C LF-HF Gain calibration:
1. Connect Ch. C of the Test Tool to the 5502A. See
The display must show step
CL 0694
(Pos C Fast). If it does not, then press
1
or
2
to select the first calibration step in
,
Input C LF-HF Gain
.
Figure 15. 5520A Scope Output to Test Tool Input C
2. Set the 5502A SCOPE output to source the signal required for the first calibration point in
Input C LF-HF Gain
.
3. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
4. Press
3
to start the calibration.
A
D
C
B
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
PM9091
PM9585 - 50
:
USE 50
:
TERMINATION
4 CHANNELS
For 190-502 and -504
use only TRM50!