190 Series III
Service Manual
48
3. Set the 5502A to supply a DC voltage to the first calibration point in
Multimeter Volt Gain
.
4. Set the 5502A to operate (OPR).
5. Press
3
to start the calibration.
Wait until the display shows calibration status
:READY
.
6. Press
2
to select the next calibration step, set the 5502A to the next calibration
point, and start the calibration. Continue through all calibration points of
Multimeter Volt
Gain
until the display shows the status
CL 0844: READY
.
Set the 5502A to Standby, and continue at the
section.
Multimeter Ohm Gain
Note
The adjustment steps for the meter section are only for the models 190-062, 190-102,
190-202, 190M-2, and 190-502.
To do the DMM Ohm Gain calibration:
1. Press
2
to select the
CL 0910
calibration adjustment step in
,
Multimeter
Ohm Gain
.
2. Connect the Test Tool to the 5502A. See
. Notice that the sense leads must be
connected directly to the Test Tool inputs.
Figure 20. Four-Wire Ohms Calibration Connections
3. Set the 5502A to the first test point in
Multimeter Ohm Gain
. Use the 5502A “COMP 2 wire”
mode for the calibration adjustments up to and including 100 k
Ω
. For the higher values, the
5502A will turn off the “COMP 2 wire” mode.
NORMAL
SCOPE
FLUKE 5502A CALIBRATOR
USE STACKABLE TEST LEADS
AS SUPPLIED WITH CALIBRATOR
A
B
METER
2 CHANN. + METER