Theory of Operation
Functional Block Description
2
2-7
A1 Main PCA Block Description
2-3.
The A1 Main pca description is divided into sections for each primary pca function as
described below.
Power Supply
2-4.
The Power Supply functional block (Figure 2-3) provides voltages required by the
outguard digital circuitry: +4.9V dc (Vcc); the vacuum-fluorescent display: -30V dc and
filament voltage of 5.4V ac; the inguard circuitry: +5.2V dc (Vdd), +5.6V dc (Vddr), and
-5.2V dc; and RS-232 interface voltage: -5.0V dc (Vee).
Within the power supply, the raw dc supply converts 107 to 264V ac line voltage into a
dc level and applies it to the power switch, and/or the 9 to 16V dc input is applied to the
power switch. The 5V Switcher (A1U9, A1U28) converts the dc from the power switch
into 4.9V +/-0.05V dc, which is used by the Inverter (A1U22, A1U23) in generating the
above-mentioned outputs. A Power Fail Detector provides a power supply status signal
to the Microprocessor in the Digital Kernel.
Within the Ethernet interface (A1U16, A1U32) there is an inverter module that provides
an isolated -9V dc supply for the 10BASE2 transceiver. The inverter module is powered
from the 4.9V dc (Vcc) supply. There is also a small power supply that provides a
programming voltage (Vpp) for the FLASH EPROM device on the outguard digital
kernel.
5V Switcher
Inverter
Regulator
Regulator
Regulator
Regulator
Regulator
Power
Switch
107 to 264
V ac In
9 to 16
V dc In
+4.9V dc (Vcc)
5.4V ac (display)
-30V dc (display)
+5.2V dc (Vdd)
+5.6V dc (Vddr)
-5.2V dc (Vss)
-5.0V dc (Vee)
Figure 2-3. Power Supply Block Diagram
Digital Kernel
2-5.
The Digital Kernel functional block is responsible for the coordination of all activities
within the instrument. This block requires voltages from the Power Supply and signals
from the Power-on Reset circuit.
Specifically, the Digital Kernel microprocessor (A1U1) performs the following
functions:
•
Executes the instructions stored in FLASH EPROM (A1U21).
•
Stores instrument calibration data in FLASH EPROM.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...