Performance Testing and Calibration
Calibration
4
4-37
Table 4-4. Manual Calibration Command Responses
Response
Description
=>
indicates the command was executed correctly.
!>
indicates that a device-dependent error was generated and the calibration step could not be
executed. Verify that the input to NetDAQ channel 1 is the correct value and polarity, is
connected correctly, and the 5700A is in the Operate mode.
?>
indicates the command was entered with incorrect syntax, for example, misspelling the
command. Reenter the command using the correct syntax.
Manual VDC Calibration Procedure
4-43.
The VDC calibration procedure calculates gain and offset calibration constants for all of
the VDC ranges. The 750 mV range is not user accessible, but is used to measure the
reference junction voltage and must be calibrated.
Complete the following procedure to manually calibrate the VDC function.
1.
If you have not already done so, complete the procedure “Preparing for Calibration”
earlier in this chapter.
2.
Connect the instrument and 5700A Calibrator as shown in Figure 4-7.
3.
Complete the sequence of manual steps shown in Table 4-5. Measuring channel 1
provides the full scale reading, and measuring the short on channel 2 determines the
zero offset error. Both readings are accomplished automatically with the
CAL_STEP? command.
Table 4-5. Manual VDC Calibration
Command
Response
Action
CAL 1
=>
Puts NetDAQ in VDC calibration.
CAL_REF?
+90.0000E-3
5700A - Source 90 mV dc.
CAL_STEP?
+90.0000E-3
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
+300.000E-3
5700A - Source 300 mV dc.
CAL_STEP?
+300.000E-3
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
+750.000E-3
5700A - Source 750 mV dc.
CAL_STEP?
+750.000E-3
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
+3.0
5700A - Source 3V dc.
CAL_STEP?
+3.0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
+30.0000E+0
5700A - Source 30V dc.
CAL_STEP?
+30.0000E+0
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
CAL_REF?
+300.000E+0 (2640A)
+50.0000E+0 (2645A)
5700A - 300V dc (2640A).
5700A - 50V dc (2645A).
CAL_STEP?
+300.000E+0 (2640A)
+50.0000E+0 (2645A)
NetDAQ computes the calibration constant and returns
the calibrated reading. Allow several seconds.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...