Diagnostic Testing and Troubleshooting
Troubleshooting the Instrument
5
5-21
Table 5-10. Relating Selftest Errors to Instrument Problems (cont)
Error
Code
Error Code
Description
Suspect
Assembly
Error Code
Discussion
4
Display test failure.
A2 Display
PCA
[Note 1]
Background The A1U1 microprocessor requests the A2
Display PCA to run a self-test and report the results.
Failure For this error to occur, the A2 Display PCA is
able to communicate with A1U1 microprocessor but the
Display self-test is “fail.”
Correction Almost certainly the A2 Display PCA has a
problem. Refer to paragraph “A2 Display PCA
Troubleshooting.”
5
Display not
responding.
A2 Display
PCA
[Note 1]
Background The A1U1 microprocessor requests the A2
Display PCA to run a self-test and report the results.
Failure For this error to occur, the A2 Display PCA did
not respond to the A1U1 microprocessor request to run a
self-test. Neither “pass” nor “fail” was reported as if the A2
Display PCA is dead or missing.
Correction Check the ribbon-cable connection between
A1J2 and A2J1, or the power supply voltages to the A2
Display PCA. Refer to paragraph “A2 Display PCA
Troubleshooting.” The A1U1 microprocessor could be
damaged for the interface with the A2 Display PCA.
6
Calibration
constants corrupt.
A1 Main
PCA
Background The A1U21 FLASH memory device is
divided into sections. One of these sections is the
memory for the calibration constants. Although the
calibration constants are used by the A/D converter on
the outguard A3 A/D Converter PCA, they are stored on
the inguard A1 Main PCA because the Flash memory on
the A3 A/D Converter cannot be programmed while the
instrument is operating, i.e., programmed with calibration
constants.
Failure For this error to occur, the calibration constants
stored in A1U21 have become corrupted. One possibility
is that you started a calibration routine and then didn’t
complete it. When you start a calibration routine, it sets a
“start” flag; when you complete a calibration routine, it
sets a “finish” flag. If the software detects a start flag and
no finish flat, you will receive this error.
Correction If the only problem is calibration values,
complete the calibration procedures in Chapter 4. One
other possibility is a A1U21 device failure.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...