![Fluke NetDAQ 2640A Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/fluke/netdaq-2640a/netdaq-2640a_service-manual_2310217019.webp)
Introduction and Specification
Introduction
1
1-3
Introduction
1-1.
This Service Manual supports performance testing, calibration, servicing, and
maintenance of the 2640A NetDAQ
™
and 2645A NetDAQ networked data acquisition
units (Figure 1-1). NetDAQ networked data acquisition units are 20-channel front ends
that operate in conjunction with a host computer to form a networked data acquisition
system. The host computer and instruments are interconnected using an Ethernet
network, and the host computer runs the NetDAQ Logger for Windows application to
provide an operating environment for the instruments, including testing and calibration.
The 2640A and 2645A networked data acquisition units are identical in operation and
appearance, and vary only in emphasis: The 2640A emphasizes precision and supports
up to 100 measurements per second, with 5 ½ digits of resolution, .02% accuracy, and
150-volt common mode voltage (300 volts on channels 1 and 11), while the 2645A
emphasizes increased measurement speed supporting up to 1000 measurements per
second, with 4 ½ digits of resolution, .04% accuracy, and 50-volt common mode voltage.
Refer to Table 1-1 for a summary of instrument specifications. For complete instrument
specifications, see “Specifications” later in this chapter.
The instruments measure dc volts, ac volts, ohms, temperature, frequency, and dc
current. Temperature measurements use thermocouples or resistance-temperature
detectors (RTDs). Refer to Table 1-2 for a summary of instrument measurement
capabilities. In addition, there are eight digital input/output lines, one totalizing input,
one external trigger input, one trigger output, and one master alarm output. The
instruments can be ac or dc powered. An RS-232 serial port is supplied for servicing and
maintenance procedures.
The term "instrument" is used in this manual to refer to both units. The model number
(2640A or 2645A) is used when discussing characteristics unique to one instrument.
Instrument assemblies are identical except for the A3 Analog/Digital Converter printed
circuit assembly (pca), which is specific to the 2640A (mechanical switching for
measurement signals) and 2645A (solid-state switching for measurement signals).
The instrument is designed for bench-top, field service, and system applications. A dual
vacuum-fluorescent display uses combinations of alphanumeric characters and
descriptive annunciators to provide prompting and measurement information during
setup and operation modes. Some features provided by the instrument are listed in Table
1-3. For additional information regarding instrument features and capabilities, refer to
the NetDAQ Users Manual (PN 942623).
NetDAQ
NETWORKED DATA ACQUISITION UNIT
COMM
DIO
MON
ENTER
CAL
ENABLE
REM SCAN
MON
V DC
CH
Figure 1-1. 2640A/2645A NetDAQ Networked Data Acquisition Units
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...