All methods use the instrument RS-232 serial interface. Procedures cannot be performed
over the network interface or from the instrument front panel.
Preparing for Calibration
4-32.
Regardless of the method you are using for calibration (automatic, semiautomatic, or
manual), the preparation for calibration is identical. Complete the following procedure to
prepare for calibration.
1.
Connect the instrument RS-232 serial port to a host computer COM serial port as
shown in Figure 4-5. You can connect two instruments as shown, although only one
is calibrated at a time. The “RS” cables used are standard null-modem cables
(reversed transmit and receive lines) that may be ordered from Fluke.
2.
Wire the Universal Input Module so the high and low inputs to channels 1 and 11 are
externally available. In addition, a four-wire short must be applied to channels 2 and
12. (See Figure 4-6.) Plug the module into the instrument.
3.
Power the instrument and allow at least a 30-minute warmup period. The instrument
must be stabilized in an environment with an ambient temperature of 22 to 24
°
C and
relative humidity of less than 70%.
4.
Using the buttons at the instrument front panel, set the desired baud rate for the
instrument RS-232 port (see below). This is the only selectable RS-232 parameter.
The other instrument RS-232 parameters are fixed: data bits=8; stops bits=1;
parity=none; echo=none; flow control=xon/xoff.
COMM
Press the COMM key to review the baud rate, or press and hold the COMM key
for 3 seconds to set the baud rate (the SET annunciator lights).
Press the up/down arrow keys until rS232 is shown in the primary display
(comm is displayed in the secondary display).
ENTER
Press the ENTER key. bAud is displayed in the secondary display and the
current baud rate in the primary display.
Press the up/down arrow keys to select the desired number baud rate:
4800, 9600, 19200, or 38400. The factory default value is 9600 baud.
ENTER
Press the ENTER key to exit. (Pressing any other function key will cancel set
operations.)
5.
Activate the calibration mode at the instrument by pressing and holding the
instrument front panel CAL Enable button for approximately 4 seconds. Release the
button after instrument beeps and CAL is shown on the primary display.
NOTE
The CAL Enable button is located on the right side of the display beneath a
calibration seal. Press this button with a blunt-tipped object. Do not press
CAL ENABLE unless you intend to calibrate the instrument. If you have
activated Calibration and wish to exit, press CAL ENABLE momentarily a
second time, or turn the instrument off.
Содержание NetDAQ 2640A
Страница 14: ...NetDAQ Service Manual x...
Страница 46: ...NetDAQ Service Manual 2 4...
Страница 106: ...NetDAQ Service Manual 2 64...
Страница 108: ...NetDAQ Service Manual 3 2...
Страница 164: ...NetDAQ Service Manual 4 42...
Страница 206: ...NetDAQ Service Manual 6 2...
Страница 218: ...NetDAQ Service Manual 6 14 2645A 1601 Figure 6 2 A1 Main PCA Assembly...
Страница 220: ...NetDAQ Service Manual 6 16 2620A 1601 Figure 6 3 A2 Display PCA Assembly...
Страница 225: ...List of Replaceable Parts Parts Lists 6 6 21 2640A 1603 Figure 6 4 2640A A3 A D Converter PCA Assembly...
Страница 230: ...NetDAQ Service Manual 6 26 2645A 1603 Figure 6 5 2645A A3 A D Converter PCA Assembly...
Страница 234: ...NetDAQ Service Manual 7 2...
Страница 242: ...2640A 2645A Service Manual 7 10 Figure 7 2 A2 Display PCA Assembly 2620A 1602...
Страница 243: ...Schematic Diagrams 7 7 11 Figure 7 2 A2 Display PCA Assembly cont 2620A 1002...
Страница 244: ...2640A 2645A Service Manual 7 12 2640A 1603 Figure 7 3 2640A A3 A D Converter PCA Assembly 2640A 1603...
Страница 251: ...Schematic Diagrams 7 7 19 2645A 1603 Figure 7 4 2645A A3 A D Converter PCA Assembly 2645A 1603...
Страница 258: ...2640A 2645A Service Manual 7 26 Figure 7 5 A4 Analog Input PCA Assembly 2620A 1604...
Страница 260: ...2640A 2645A Service Manual 7 28...