a3003 FC/a3004 FC
Calibration Manual
18
2000 mA Clamp Calibration Adjustment
Do the balance adjustment and then do the calibration adjustment.
To do the balance adjustment:
1.
Follow the instructions to enter Maintenance mode (see
) and then:
a.
Wait 20 seconds for the Meter to warm up.
b.
Push
until
JC
shows on the display.
c.
b-UP
shows on the display.
2.
For each value in Table 7:
a.
Apply the Calibrator Output value to the Meter.
b.
Wait 4 seconds for the signal to stabilize.
c.
Push
to confirm the calibration step.
3.
On the 5522A, set the 12 A dc output to the 50-turn 1 kA coil and place the clamp on it.
4.
Push
, until
b-UP
shows, then move the jaw and locate the coil center in the top area of the jaw.
5.
Push
, until
b-do
shows, then move the jaw and locate the coil center in the bottom area of the jaw.
6.
Repeat the previous two steps until
0 A
shows.
When 0 A shows on the display the balance adjustment passes. Continue with the calibration.
To do the calibration adjustment:
1.
Move the jaw to the 3 kA coil.
2.
Power on all the equipment and set the parameters:
•
On the 5522A, put in the Standby state with 0
A output.
•
On the 52120A put in Current Mode input, with the 120 A range. Put the 52120A in Local Control mode.
3.
Push
on the Meter.
4.
For each value in Table 8:
a.
Apply the Calibrator Output value to the Meter.
b.
Wait 4 seconds for the signal to stabilize.
c.
Push
to confirm the calibration step.
If
Err
shows on the Meter when you push
, the internal measurement is working and the result is not ready.
Wait for 1 minute.
If
out
shows on the Meter, the internal measurement is out of range. Check the input signal and correct it.
Table 8. Calibration Specifications 2000 A Clamp
Display
Output to 3KA coil
From 5522A
[1]
0A
0 A
0 mA
C-01
10 A
0.4 mA
C-02
1000 A
40 mA
C-03
2000 A
80 mA
H-0A
0 A
0 mA
C-04
-10 A
-0.4 mA
C-05
-1000 A
-40 mA
C-06
-2000 A
-80 mA
[1]: Manually adjust the current signal from the 5522A to drive the high current from 52120A.