Wireless DC Clamp
Maintenance Mode
7
Maintenance Mode
Use Maintenance mode to set parameters on the Meter. Some options require a password. Maintenance mode
options are:
•
Full-screen mode
•
Revision information (rxxx, where xxx is the revision)
•
Poff
•
Loff
•
FC
•
CAL
(Calibration mode)
•
JC
(Jaw Calibration)
•
TEST
(Verification mode)
Start Maintenance Mode
To enter Maintenance mode:
1.
With the Meter off, push and hold
.
2.
Push
while you keep
pushed until all of the display segments are shown.
3.
Release
and
, to enter the maintenance mode.
4.
Push
until the appropriate option shows on the display.
Enter the Password
The CAL and JC options require a password.
To enter the password:
1.
Push
.
The Meter shows the CAL counts, for example
n002
.
2.
Push
to show
????
.
The first digit
flashes
3.
Push
and enter the first digit of the password. (Default is 1234).
4.
Push
and enter the next digit of the password, until all digits are entered.
5.
Push
to confirm the input.
When the correct password is entered the Meter shows
C-01.
When the incorrect password is entered
????
shows and you must re-enter the password.
Change the Password
Note
If you forget the password after you change it, you can reset it to 1234. See
To change the password:
1.
Do steps 1 through 4 in the
section.
Do not confirm the entry.
2.
Push
to show ---- on the display. The first digit flashes.
3.
Push
and enter the first digit of the new password.
4.
Push
to accept the first digit, then enter the next digit, continue until all digits are entered.
5.
Push
to confirm the changes.
If the Meter has been calibrated, the Meter goes to normal measurement mode. Otherwise, the Meter shows
donE
.
Содержание A3003 FC
Страница 4: ...ii a3003 a3004 FC Calibration Manual...