Wireless DC Clamp
Calibration
15
2000 mA Clamp Balance Test
Do the balance test after calibration. The balance test evaluates the clamp measurement error when the clamp is in
different positions. If any position fails the balance test, a balance adjustment is required. See
To do the balance test:
1.
Make sure you have the required equipment. See Table 2.
2.
Connect the equipment. See Figure 6.
3.
Power on all the equipment.
4.
On the 5522A, set the 9.0 A dc output to the 50-turn 1 kA coil and place the clamp on it.
5.
Push
until
b-UP
shows, then move the jaw and locate the coil center in the top area of the jaw. Hold for
5 seconds.
6.
Push
until
b-do
shows, then move the jaw and locate the coil center in the bottom area of the jaw. Hold for
5 seconds.
7.
Record the reading.
The reading difference between the top and bottom position must be below 1.0 A (10 digits.)
Calibration
Use the calibration procedures to adjust the Product so that the values shown on the Meter correspond as closely as
possible with the actual measured values. The Product measures the applied reference source, calculates correction
factors, and stores the correction factors in nonvolatile memory.
Should the Product fail any of the performance tests, do the calibration adjustment procedure.
Note
The first calibration of the Product is available from the Fluke factory only.
Meter Calibration Adjustment
Before you can adjust the Meter calibration, you must enter your password in the Maintenance Mode.
1.
Follow the instructions to enter Maintenance mode (see
) and then:
a.
Push
until
CAL
shows on the display.
b.
Follow the steps in
until
C-01
shows on the display.
2.
For each value in Table 6:
a.
Apply the Calibrator Output value to the Meter.
b.
Push
to confirm the calibration step.
If the input signal does not satisfy the calibration requirement, the Meter shows
Err
. If the signal is not stable, you
must push
several times to confirm the calibration.
After you confirm the calibration, the Meter goes to the subsequent calibration step. When the last calibration
point is recorded,
End
shows on the display. See Figure 7.
Note
After you push
wait for the calibration step number to advance before you change
the calibrator source. Some adjustment steps can take several seconds to complete
before the Meter goes to the subsequent step. Set the Calibrator to Standby after you
complete adjustment of each function.
Note
While the calibration adjustment points are shown in Table
the Meter shows the
necessary inputs. For each step, push
to see the necessary dc reference signal
amplitude. Use the connections shown in Figure 8.